IEC 60601-1:

Translating 60601 Requirements into Results

Translating 60601 Requirements into Results

Planning is Vital

Medical electrical equipment must comply with IEC 60601-1 in order to conform to the legal requirements of the Medical Device Directives. This applies to existing devices on the market and to new devices currently under development.

New Design Input Requirements

  • Manufacturers should specify the latest edition of 60601-1 as a new regulatory requirement within the Design Input Requirements
  • The manufacturer should also ensure that the design team is aware that the standard will have potential implications not just for the device design, but may also trigger requirements for labelling, instructions for use (IFUs), and risk management plans

Intended Use and Application

  • Document the intended purpose of the device, together with any additional parts or accessories necessary to meet the intended use
  • Communicate this data to all the design team members, including the appointed regulatory personnel on the team
  • Plan this activity early in the design/development cycle to enable a thorough review and correct application of the standard by the design team
  • Ensure that the documentation thoroughly describes how the equipment or device is intended to be applied in practice and has specified the intended users and the application environment
  • Include this documentation within the design history and risk management files

Design Planning

  • Specify the evaluation of the relevant IEC as an early phase activity in the design plan
  • Apart from product design implications, risk management records will need to demonstrate evidence of conformance to the requirements
  • In addition, product labels may be impacted, and instructions for use may need to be changed as a consequence of the evaluation process
  • The design plan should be kept up to date with new tasks and the required technical and regulatory resource requirements
IEC 60601, Medical Electrical Equipment Testing

IEC 60601, Medical Electrical Equipment Testing

General, Particular, and Collateral Standards

  • We include detail on particular and collateral standards – should we insert a bullet on general standards?
  • Particular standards may modify, replace, or delete requirements contained in the General Standard or Collateral Standards
  • Collateral standards complement 60601-1 and specify general requirements for safety applicable to a group of medical electrical equipment and/or a specific characteristic of all medical electrical equipment, not fully addressed in the General Standard

Design Verification Planning and Reports

  • Design verification evidence will be necessary to demonstrate conformance to requirements
  • All planned design verification activities should be captured within documented and approved design verification protocols
  • Corresponding design verification evidence may take the form of tests and documented evaluations
  • Some may be focused solely on electrical, safety, or mechanical requirements. Others may be in the form of documented inspection of risk management records and documented inspections of device labelling versus the requirements of the standards

Relationships between General, Particular & Collateral Standards

Risk Management File

  • A risk management file is created in accordance with ISO14971
  • The risk management plan should identify the review of the IEC 60601 series of standards as a risk management activity
  • The risk analysis, risk evaluation, and risk control phases should methodically address each applicable requirement
  • The risk management report should confirm that all applicable requirements have in fact been addressed

Prepare with the End Results in Mind

These are some exemplary questions to consider as part of preparing for a successful outcome and to understand the scope of the task ahead:

  • If a regulator samples clauses from applicable IEC 60601 standards, what documented compliance evidence will exist within Design History File records?
  • If the device labelling and instructions for use are examined, will they have adequately addressed all of the applicable requirements within the standards?
  • What documented evidence will demonstrate that risk management records have met the requirements of ISO14971 and the 60601 series requirements?

The answers to the above will be derived by a methodical review of the requirements, discrete activity planning within the design control process, and by appointing the experienced regulatory and design personnel necessary to complete the tasks.